Elevated Privileges Vulnerability in Smiths-Medical CADD-Solis Medication Safety Software

Elevated Privileges Vulnerability in Smiths-Medical CADD-Solis Medication Safety Software

CVE-2016-8355 · HIGH Severity

AV:N/AC:L/AU:S/C:C/I:C/A:C

An issue was discovered in Smiths-Medical CADD-Solis Medication Safety Software, Version 1.0; 2.0; 3.0; and 3.1. CADD-Solis Medication Safety Software grants an authenticated user elevated privileges on the SQL database, which would allow an authenticated user to modify drug libraries, add and delete users, and change user permissions. According to Smiths-Medical, physical access to the pump is required to install drug library updates.